125
A II. ADJUSTMENT
MARGINS TABLES
Main voltage measurement points
CIIM0015 Power supply board
TP1 - 6V [5.6-5.8] V
TP2 - 12V [11.7 -11.9] V
TP3 - 5V [4.9-5.2] V
Sp150 [23.5-24.5] V
CIIM0026 Microprocessor board
TP60 - Vref motor Z [13 -15] mV
TP61 - Vref motor Y [13 -15] mV
TP56 - Vref motor X [13 -15] mV
TP57 - Vref pump motor [13 -15] mV
TP58 - Vref lter motor [13 -15] mV
TP59 - Vref rotor motor [13 -15] mV
Main adjustment values of the analyzer
Tolerances
Operating arm positioning
X axis loss of steps
Loss of steps [-3,3]
Y axis loss of steps
Loss of steps [-3,3]
Z axis loss of steps
Loss of steps [-3,3]
Loss of steps rotor motor
Loss of steps [-3,3]
Loss of steps lter drum motor
Loss of steps [-3,3]
Loss of steps pump motor
Loss of steps [-4,4]
XY position adjustment
X origin [10,125]
Y origin [240,290]
X racks tray [620,750]
Y racks tray [-20,20]
X washing station [330,450]
Y washing station [70,120]
X rotor [60,170]
Y rotor [1030,1060]
Maximum Z [1090,1155]
Self-centering of needle
X Offset [-4,4]
Y Offset [-4,4]
Bottle level control sensitivity
Water empty [200,255]
Water full [140,230]
Water calculated [170,240]
Waste empty [200,255]
Waste full [170,245]
Waste calculated [200,245]
Level detection sensitivity
Paediatric rack [45,90]
13/15mm tube sample rack [45,90]
Reagent rack [25,45]
Positioning of the dispensing point
Dispensing rotor [90,120]
Fine rotor X [35,200]
Positioning of rotor in optic
Positioning [-15,0]
Needle thermostatation
Set point temperature [35,45]
Rotor thermostatation
Set point temperature [36,37.5]
Filter wheel positioning
Positioning [-5,5]
Integration times
F1 340 nm [190,350]
F2 405 nm [40,130]
F3 505 nm [40,130]
F4 535 nm [40,130]
F5 560 nm [40,130]
F6 600 nm [40,130]
F7 635 nm [40,130]
F8 670 nm [50,130]
Number of counts
F1 340 nm [750000,960000]
F2 405 nm [840000,940000]
F3 505 nm [840000,940000]
F8 670 nm [840000,940000]
Darkness counts
F1 340 nm [3700,4300]
F2 405 nm [3700,4300]
F3 505 nm [3700,4300]
F8 670 nm [3700,4300]
SMF repeatability (Noise)
Dev. Est. NC. F0 Covered 55
Dev. Est. Abs. F1 340 nm 0.0004
Dev. Est. Abs. F2 405 nm 0.0006
Dev. Est. Abs. F3 505 nm 0.0004
Dev. Est. Abs. F8 670 nm 0.0007
Stability at 505 nm
Dev. Est. Abs. F3 505 nm 0.0008
Max. NC [840000,940000]
Min. NC [840000,940000]
Max NC / Min NC 1,006
Repeatability MF (Repeatability)
Dev. Est. Abs. F1 340 nm 0.0008
Dev. Est. Abs. F2 405 nm 0.0006
Dev. Est. Abs. F3 505 nm 0.0005
Dev. Est. Abs. F4 535 nm 0.0005
Dev. Est. Abs. F5 560 nm 0.0005
Dev. Est. Abs. F6 600 nm 0.0005
Dev. Est. Abs. F7 635 nm 0.0005
Dev. Est. Abs. F8 670 nm 0.0005