EasyManua.ls Logo

biosystems A15 - Changing the Arm Hose

biosystems A15
131 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
99
b) Disconnect the connector (1) that goes to the board that goes to the needle unit and cut all the anges required.
c) Disconnect the Teon dispensing pipe (2).
d) Remove the cover that covers the X carriage interconnection board (3). Cut the anges that hold the at bands
and disconnect them from the board.
e) Unscrew the four studs that fasten the two X guide axes and which are located on the arm supports.
f) Pull the two guide axes as shown in the gure.
5.1.2.2. Changing the arm hose
a) Remove all the analyzer housing.
b) Disconnect the connector (1) that goes to the board that goes to the needle unit and cut all the anges required.
c) Disconnect the Teon pipe at both ends (needle and electrovalve) (2). Cut all the anges that guide the pipe to the
electrovalve.
d) Remove the cover (3). To do so, open the electronics box at the rear and remove the fastening clips. Disconnect
the connector that goes to the CPU board and the earth connection (6).
e) Unscrew the ange screw (4).
f) Replace the hose, connect and replace the anges.
1
3
4
2

Table of Contents