Service manual
108
5.1.4. Reactions rotor and reading
5.1.4.1. Changing the rotor temperature probe
a) Remove the upper casing.
b) Disconnect the electrical hose from the interconnection board.
c) Remove the thermal insulation from the temperature probe.
d) Unscrew the probe (2).
e) Clean the thermal silicone form the housing and put fresh thermal silicone on the end of the new probe.
f) Fit the new probe.