The programme automatically saves a copy of the adjustments and cycles read in a le. This le is located in
the following folder:
c:\Program les\A15 Service\Adjustments\
When a physical element of the analyser has to be changed, e.g. the Z axis belt, the counter must be reset
to zero for it to correspond to the number of cycles actually stored in the analyser. To perform this operation,
select the box of the element that is to be initialised and enter the number of cycles in the enabled box. Then
press the Load cycles button.
Using the Load adjustments button, this screen also enables the technician to select an adjustments le and load
it in the analyser. When the adjustments are loaded, the cycles are also loaded. Perform this operation when
a CPU board has to be changed. This avoids having to completely readjust the analyser; only the following
sections will have to be readjusted:
• Scales
• Level detection sensitivity
• Needle thermostatation
• Rotor thermostatation
4.4.8 Change the rotor type
In this screen the type of rotor is introduced. Each rotor comes labelled with a letter in its top part. Select in
this screen the type of rotor to use. For rotors marked with A letter, only select the letter. For the rotors marked
with other letters, select OTHERS and then introduce the light path that will come it within the box of rotors or
of the distributor.