EasyManuals Logo

biosystems A15 User Manual

biosystems A15
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
79
4.3.8. Level control scales test
This screen makes it possible to check the functioning of the level control scales of the waste and distilled water
containers. The technician must select which scales he wishes to check and place a certain amount of liquid
in the corresponding container. On pressing the Test button, the screen shows the level of liquid measured by
the analyzer(only 0 % and 100%)
4.3.9. Covers detection test
This test makes it possible to check the functioning of the different detectors incorporated in the analyzer.
- Open detector of the general cover of the analyzer.
Rotor cover presence detector.
The technician can manipulate the corresponding components, for example, open and close the cover of the
analyzer and the screen shows the state of the detectors in each case.

Table of Contents