Service manual
4. SERVICE PROGRAM
The service program is used for the adjustment, checking and maintenance of the different components of the
analyzer. It is not supplied with the instrument, it is supplied to authorised technical services only. The personal
of the Technical Service must install it on the user’s computer in order to carry out the service requirements.
Once the tasks have nalised, the program must be uninstalled. To install the program, follow the instructions
on the installation CD ROM called Service. The original password for using this program is A15. The password
can be changed from the service program itself. If the service personnel forget the password, the original pas-
sword can be reinstalled by deleting the hidden le code.A15 from the application directory and relaunching
the program. Once the password has been introduced, the analyzer serial number is given and the name of
the operator is requested (by default Operator1). Press the Accept button and the main program appears. The
different functions of the service program are classied in the following categories:
- Adjustments: These make it possible to make different parameter adjustments required for the correct
functioning of the analyzer.
- Tests: Tests for checking the functionality of the a na ly zer.
- Utilities: Different technical utilities, such as, for example, washing or priming the dispensing sys tem
or changing an optical lter.
- Registers:This enables the management of past adjustments, tests, incidences, repairs and maintenance
of the instrument.
- Monitor:These enable the low level communication with the analyzer to load new versions of the program
in the ash memory of the analyzer (rmware) or to consult the internal parameters of the instrument.
An emergency stop button (STOP on a red background) will be accessible at all times, and when pressed, it
switches off the analyzer and closes the application quickly.
4.1 Initialising the analyser
To initialise the analyser in service mode, rst launch the A15 Service application. The program rst of all re-
quests a user or technician ID to be used in the program. Depending on the type of user identied, access to
the different parts of the program will be allowed or denied. The following screen appears:
For full access, enter the following codes:
Name (login): SAT
Password: A15