60
Service manual
4.2. ADJUSTMENTS
These make it possible to make different parameter adjustments required for the correct functioning of the
analyzer. All the values to be adjusted have certain limited ranges, indicated by the service program. These
values are also given in an appendix at the back of this manual. If, after varying any of the parameters within
its permitted range, the analyzer is not tuned up, it indicates that the corresponding system is broken and in
need of repair.
4.2.1. Adjustment of the needle thermostatation system
This screen adjusts the needle thermostatation in such a way that the dispensing temperature of the reactions