81
can be cancelled at any time.
Once the test has been launched, the screen provides regular information about the current status of the pro-
cess. If an error occurs during the process, the test ends and the screen displays a message indicating the
element causing the error.
Partial stressing of the elements of the analyser is possible. The following elements can be stressed partially:
• X axis
• Y axis
• Z axis
• Reactions rotor
• Filter wheel
• Dispensation pump
• Membrane pumps
• 2-way electrvalve
• 3-way electriovalve
4.3.12. Photometry tool
This option is used with the Photometry tool (AC15222). It is used for automate the reading process of the tool.
To use the photometry tool folow the next stetp:
1. Switch on the analyzer.
2. Run the service programme. (The older version to use the tool is the 4.1)
3. Left the analyzer switch on initialized with the service programme for 20 minutes, to preheat the lamp.
4. Select the test menu and Photometry tool option.
5. Press the Load Parameters button.
6. Insert the CD-ROM and select the le ReferenciaUtilFotometria.bin. Push Accept button.
7. Insert the tool (1) in the place of the rotor.
8. Press the Read ABS button.
9. Press the Report button to print the results report.