Service manual
4.3.10. PC-Analyzer communications channel test
On pressing the Test button, the computer attempts to establish communication with the analyzer. The program
tells the technician if it has been possible or not.
The technician can select Automatic Conguration or Manual Conguration. In the case of the latter, he can
dene the Port and the Speed.
4.3.11. Global stress mode of the analyzer
This test makes it possible to continually reproduce work cycles of the analyzer similar to those made during
the preparation and reading of reactions in a normal working routine, but dispensing at the washing station
instead of the rotor. It is necessary for the dispensing system to be primed so that the piston does not function
dry. All the racks must be removed from the racks tray. This test can be made without the covers and housing
of the analyzer. The technician can program the number of cycles he wishes (1 cycle = 15 seconds). The test

Table of Contents