Change in the versions of user program
Date version Changes
04 /04/06 3.2.1 First release version
20/07/06 3.2.2 New button in the tools menu, reset the historical base
line
Fix the options of the initial and nal wash. Delete the menu to
choose the parameter of initial and nal wash. Add a maintenance
guide in the help menu.
14/11/06 3.3.0 Improve and optimitation of the execution times of the application.
Corrected the application errors that showed the message «not res-
pond»
05/03/07 3.3.1 Korean language included
Change of some test adapters
Correction of low frequency errors
21/03/07 3.3.2 Added the Korean language. Correction of Run Time errors.
14/06/07 4.0.0 Added the Romanian and indonesian languages
Added the substrate depletion control function in kinetic test
Added the prozone control function in turbidimetric test
Added the correlation of result per test with other analysers
Added graphical representation of results
Change the reagent Blank use in calibration factor calculation
Added a new patient results report
Added a new document explaining the Remarks messages
17/07/09 4.1 New function of LIMS online communication.
New function of the calibration plot: when a concnetration value is out
of the plot it can repeat with a postdilution factor.
Change in the versions of rmware
Date version Change
04/04/06 2.90 First release version
09/05/06 2.92 Improvements in the generation of the pattern of encoder
18/12/06 3.10 In order to avoid the splashes in the dispensation of the predilution,
the dispensation speed have been reduced.
The sensitivity adjustment have been split in 3 diferents ad-
justments, one for each type of rack.
Correction of the bireagent with 10 minuts of reading time and with
programmed contamination.
6/02/07 3.12 Increase the stabilization time of the commutation of 3 way electro-
valve.
07/03/07 3.14 Correction: when shut-down is made and next the warming-up wi-
thout disconnecting the analyzer the termostatización of the probe
does not activate.
15/05/07 3.20 Added an autodiagnosis system for internal use.
29/10/07 3.24 Correction of the transition state from shutdown to slee-
ping. Improved the swith-off of global-stress
Correction of initial washing when previously the cover was open
and close.
16/04/08 3.28 Correction of the predilution manoeuvre when tube rack is used
22/01/09 3.50 Improve the adjustment position of the racks and reagents racks.
New sensibility maps of the improved racks
17/07/09 3.61 Include both types of sensibility maps: metal maps and plastic
maps.
Correction about enter the serial number of the analyzer
Added a protection of the Shut-down process.