Unit Diagnosis / Calibration
2
SpaceStation, 2.0 gb 2- 3
Alarms and Error Codes
SpaceStation without SpaceCom
Neither alarms nor error codes are actually displayed on the
SpaceStation without SpaceCom.
SpaceStation with SpaceCom
SpaceCover comfort
Alarm Possible Cause Fault Clearance
1 SpaceComAlarm indicator (red) is on SpaceCom is still starting
2 SpaceComAlarm indicator (red) flashes A serious internal fault was detected in
the system
Carry out a device check (see „Device
Check“ ➨ S. 2 - 4)
Table 2 - 1 Alarms on the SpaceCom
Alarm Possible Cause Fault Clearance
1 One of the diodes of the status- and
alarm display line is on or flashes
A message or an alarm of a pump is dis-
played
2 The status indicator (on the right-hand
side of the On-/Off-button) flashes red
System not properly configured Check configuration
3 The status indicator (on the right-hand
side of the On-/Off-button) lights up red
A serious internal fault was detected in
the system
Carry out a device check (see „Device
Check“ ➨ S. 2 - 4)
4 The first LED in the top charge condition
display of the operator and status display
field of the SpaceCover comfort flashes
yellow
Battery module of theSpaceCover com-
fort nearly discharged (type: pre-alarm)
Connect device to voltage supply
5 The first LED in the bottom charge condi-
tion display of the operator and status
display field of the SpaceCover comfort
flashes yellow
Battery module of theSpaceCom nearly
discharged (type: pre-alarm)
Connect device to voltage supply
6 The first LED in the top charge condition
display of the operator and status display
field of the SpaceCover comfort flashes
red
Battery module of the SpaceCover com-
fort discharged
Connect device to voltage supply
7 The first LED in the bottom charge condi-
tion display of the operator and status
display field of the SpaceCover comfort
flashes red
Battery module of the SpaceCom dis-
charged
Connect device to voltage supply
Table 2 - 2 Alarms of the SpaceCover comfort (Part 1 of 2)
2.0