EasyManua.ls Logo

Braun SpaceStation - Procedural Instructions on the TSC; Visual Inspection Procedures; Electrical Safety Measurements

Braun SpaceStation
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SpaceStation, 2.0 gb 8- 1
8
8-Procedural Instructions on the TSC
Visual Inspection
1. Check the unit and accessories for cleanliness.
2. Check the unit and accessories for completeness and check
configuration.
3. Check the unit and its accessories for damage and the labels
for readability. Pay special attention to the following parts:
a) Pump suspension
b) Tightness of pump locks
c) Connectors
d) 8 tube guides
e) Mains cable
Electrical Safety
according to IEC/EN 60601-1
or VDE 0750 and VDE 0751
The values to be measured for the electrical safety are stated in
the TSC (see „Technical Safety Check (TSC)“ S. 7 - 1).
Protective Conductor Resistance
Note
The protective conductor resistance of each SpaceStation must be
measured separately. Before carrying out the measurement re-
move all interface cables to the other units.
Measure between the protective conductor of the mains cable
and the following parts:
- Ground contact in connector F1B
(middle pin)
2.0