Procedural Instructions on the TSC
8
8- 2 SpaceStation, 2.0 gb
Earth Leakage Current
Note
Due to the technical design the patient- and housing leakage cur-
rent is not to be measured on the SpaceStation without Space-
Com and without pump inserted. The earth leakage current is
always measured as it is the upper limit of leakage currents which
can be measured.
The earth leakage current of each SpaceStation must be measured
separately.
Check
1. Before carrying out the measurement remove all interface
cables to the other units.
2. Remove all pumps out of the SpaceStation.
3. Measure the earth leakage current without single fault con-
dition (NC = Normal Condition) incl. mains cable.
4. Document value.
5. Measure the earth leakage current in single fault condition
(SFC = Single Fault Condition) incl. mains cable.
6. Document value.