106 T419S Indicator Service Manual
9 UNIT Unit of Measure Type Unit
STRING
Unit of Measure
Applies to token 9 only
U
1 = Current (default)
2 = Calibration
3 = lb
4 = kg
5 = oz
6 = g
7 = lb-oz
8 = Custom Unit 1
9 = Custom Unit 2
10 = Custom Unit 3
11 = Custom Unit 4
5WST
Motion = M
Stable or no motion = S
sMotion
15 WSTAT
Weight Status
See Additional token tables on page 109
Weight Status
17 TSAT
Returns type of Tare used
See Additional token tables on page 109
Tare Type
28 HEX
Prints HEX value of a ASCII Decimal Value
Hexadecimal Applies to token 28 only.
non-optional parameter
[hXX]
0 through 255 (See ASCII chart for the Unicode UTF8 characters.)
29 NCIBYTE1
Returns NCI Status Byte 1 only
See Additional token tables on page 109
NCI Status Byte 1
30 NCIBYTE2
Returns NCI Status Byte 2 only
See Additional token tables on page 109
NCI Status Byte 2
31 NCIBYTE3
Returns NCI Status Byte 3 only
See Additional token tables on page 109
NCI Status Byte 3
32 NCIBYTE12
Returns NCI Status Bytes 1 & 2 only
See Additional token tables on page 109
NCI Status Bytes 12
33 NCIBYTE23
Returns NCI Status Bytes 2 & 3 only
See Additional token tables on page 109
NCI Status Bytes 23
34 NCIBYTE123
Returns NCI Status Bytes 1, 2 & 3
See Additional token tables on page 109
NCI Status Bytes 123
253 T.OUTS Outputs 1-64 outputs
I/O token See note on page 107
254 T.INS Inputs 1-64 inputs
309 SITE Site ID Site ID
310 SN Serial Number Serial Number Justification J
1 = Right (default)
2 = Left
311 LK Condec Units = L or K Condec Units
Width W 1 - 127 Characters
312 STAT
Condec Status = “ “ (space = stable)
“M” = motion
“O” = out of range
Condec Status
323 S127 WI-127 Status Byte 127 Status
STATUS CODES
These are the most common characters yoiu will see on a terminal:
“0” = Valid weight
“1” = Motion
“2” = Range error (Over/Underload)
“4” = A-D error
“8” = Low voltage
Combinations of these errors can also occur. (e.g., “3 = Range error (2) plus
Motion (1))
324 NAME Indicator Scale Name Scale Name
32 characters max entered in Z
tools under SYSTEM TAB
325 LOC Indicator Scale Location Scale Location
Token Token Description of Value Name Value Data Type Parameter Name Optional Parameter Available Values
T.x T.x.1 T.x.2 T.x.3

Table of Contents

Other manuals for BTEK T419S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BTEK T419S and is the answer not in the manual?

BTEK T419S Specifications

General IconGeneral
BrandBTEK
ModelT419S
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals