40
5.1 Treadmill Shock Detection Setup
Switch on the Treadmill Shock Detection by fully inserting the DC Coaxial power plug into the power
receptacle on the rear panel of the unit.
The first option of the menu should be displayed “Mode: Manual”.
If the unit already has power, you may press the “RESET” button to quit the current operation and return to
the top of the menu (“Mode: Manual”).
5.2 Incline Controller Setup
Turn the “RATE” knob counter-clockwise, all of the way to the end-stop.
Set the mode toggle switch to “STOP”.
Now, switch on the Incline Controller. The power switch is above the AC power receptacle on the rear
panel.
The front panel display should report the position of the incline knob (upper left), the incline of the
apparatus (upper right), the minimum incline rate (0.1°/s, lower left) and the mode of “STOP”.
Adjust the “INCLINE” knob so the desired position matches the apparatus’ incline position to ensure
the apparatus will not move when enabled.
Now set the mode toggle switch to “RUN”.
Adjust the “INCLINE” knob to 0.0°.