ECG, Arrhythmia, and ST Segment
142
Instructions for Use Infinity Acute Care System - Monitoring Applications SW VG1
Connecting the Lead Wires for Neonatal Monitoring
1 Insert the ECG adaptor cable (B) into the
recessed channel (A) on the side of the M540.
Orient the ECG adaptor cable so the exposed
pins face towards you as you push it firmly into
the channel.
2 Insert the spacer (C) to protect the unused
ECG-lead pins on the M540.
3 Connect the individual neonatal ECG lead wires
(E) to the neonatal ECG adaptor cable (D).
For information on applying the electrodes to the
patient, refer to the illustrations starting on
page 140.
313
A
B
C
E
D