Instructions for Use Infinity Acute Care System - Monitoring Applications SW VG1 331
Problem Solving
ARR
! The ECG
unplugged
(3)
The ECG lead wires are
disconnected from the
M540.
Check the cables and
connections; replace if
necessary.
None ST relearn ST relearn is in progress. Information only - no action
required.
(3)
In the parameter box, the parameter value is replaced by ***
Priority Message Problem Solution
Priority Message Problem Solution
!!! Asystole The indicated arrhythmia
was detected .
Check the patient and treat, if
necessary.
Some messages only appear
when the full arrhythmia option is
installed.
!!! Ventricular
Fibrillation
!!! ARR Ventricular
Tachycardia
(1, 2)
!! ARR Run
(1, 2)
!! ARR Accelerated
Idioventricular
Rhythm
(1.2)
! ARR Supraventricular
Tachycardia
(1, 2)
! ARR Couplet
(1, 2)
! ARR Bigeminy
(1,2)
! ARR Tachycardia
(1, 2)
! ARR Bradycardia
(1,2)
! ARR PAUSE
(1,2)
!! ARR artefact
(1,2)
None ARR cannot learn After 100 beats, the M540
cannot determine the
dominant normal complex
on any lead selected for
QRS processing.
Check the electrode
preparation.
Reapply electrodes, if
necessary.
NOTE
(1)
These arrhythmia events can have one of three alarm priorities : high, medium, or low.
(2)
For these arrhythmia events, you can disable the alarms.