Alarms
82
Instructions for Use Infinity Acute Care System - Monitoring Applications SW VG1
OR Mode
When the OR mode is enabled, the following
happens at the Cockpit
– Alarm messages for medium-priority and high-
priority alarms clear when the alarm condition
no longer exists.
To enable or disable this feature, see page 276.
French NFC Mode
When this mode is enabled, the following happens
at the Cockpit:
– HR alarms cannot be disabled.
– The alarm pause period cannot last longer than
3min.
– You cannot enable the cardiac bypass mode
when the French NFC mode is enabled. If the
cardiac bypass mode was enabled before
French NFC mode was turned on, the cardiac
bypass mode is disabled.
To enable or disable this feature, see page 295.
Pre-Silencing Alarms
This function allows you to concentrate on a
procedure without being interrupted by continuous
audible alarm signals arising from potential alarm
conditions. When you pre-silence alarms, the
following happens at the Cockpit:
– Visual alarms are reported as usual (see
page 76).
– The alarm message Audio paused appears in
the far-right field of the header bar along with a
countdown timer and the following symbol:
To pre-silence alarms
z Press the yellow key on the Cockpit.
Pressing the key again cancels the pre-silence
state.