Instructions for Use Infinity Acute Care System - Monitoring Applications SW VG1 355
Maintenance
Maintenance of the IACS Components
The normal inspection interval for a Cockpit is 24
months. This interval is shorter if a device with a
shorter inspection interval, that depends on the
Cockpit for its proper function, is connected to the
Cockpit.
CAUTION
This device must be inspected and serviced at
regular intervals. A record must be kept on this
preventive maintenance. We recommend
obtaining a service contract with DrägerService
through your vendor. For repairs, we recommend
that you contact DrägerService.
NOTE
The following table provides an overview of the
recommended maintenance intervals for the IACS
components (Infinity C500, C700, and PS250).
Device part Maintenance interval and tasks
NIBP air intake filter of the M540 – Exchange after two years.
– Dispose of the part with normal domestic
waste.
Internal M540 battery and internal PS250 battery – Service during regular inspections.
– Exchange at least every two years by trained
service personnel.
– Check the battery once a year. If necessary,
trained service personnel must replace it.
Inspection of all devices Exchange every two years by trained service
personnel.
Safety inspection of all devices Every year by trained service personnel according
to hospital protocol. For the scope of the safety
inspections, see page 356.
NOTE
If the NIBP air filter seems dirty or damaged,
replace it before the recommended two years.
The NIBP air filter should be replaced, if the M540
was exposed to liquid. See "Exchanging the
ambient air filter" in the Technical documentation
which is available from DrägerService.
NOTE
Do not disassemble the Ni-MH battery inside the
PS250 power supply. Aside from the required two-
year maintenance recommended for the entire
IACS, this battery requires no additional routine
maintenance.
NOTE
During high-performance use of this device, the
battery must be checked more often.