Instructions for Use Infinity Acute Care System - Monitoring Applications SW VG1 243
Cardiac Output (C.O.)
Manual Mode Measurements
If automatic measurements are not possible due to
unstable blood temperatures or other causes,
switch to the manual mode. To select the manual
C.O. mode, see page 247.
To start a measurement in the manual mode
1 Press the
C.O. Start key (A) on the
QuadHemo MPod or the Hemo4 pod/Hemo2
pod.
or
Select Procedures... from the main-menu bar
then select the C.O. tab to access the
Procedures C.O. page. Select the
C.O. Start
button on the Cockpit. Inject the saline solution
immediately.
2 Repeat steps 1 and 2 for additional
measurements.
The Procedures C.O. page (see page 245)
stores up to five C.O. averages with time
stamps. Each value panel is touch-sensitive
and allows you to include or exclude a value
from the calculation of the average. Any value
that is crossed out is excluded from the
average. If you touch the panel again, the value
re-appears and will be included in the average.
To save the C.O. average, see page 244.
340339
Quad Hemo MPod
A
A
Hemo2 Pod
338
A
Hemo4 Pod