Alarms
Auto-Setting All Alarm Limits
The auto-set function allows you to adjust alarm
limits quickly based on preset percentages listed in
the following table.
You can either auto-set:
Individual parameters (see page 96)
All parameters (see page 96)
All ST parameters (page 96)
Parameter Upper Limit Lower Limit
Ta, Tb, T1a, T1b,
Tblood
107 % of the current value 93 % of the current value
ΔT, ΔT1, PVC/min,
inCO2
Unaffected Unaffected
SpO
2 Adult/paediatric: 100% saturation
Neonate: 98 % saturation
Current value -((value)*(5 %))
ST Current value +2.0 mm Current value -2.0 mm
etCO
2 Current value +25 % Current value -20 %
All others Alarm limit that is closest to but not
greater than 25 % above the current
value of the parameter
Alarm limit that is closest to but not
greater than 20 % below the current
value of the parameter
NOTE
If the Auto-set function forces the alarm limits of a
parameter outside the allowable limit range of the
monitor, the alarm limits remain unchanged.