Problem Solving
348
Instructions for Use Infinity Acute Care System - Monitoring Applications SW VG1
Recording Status Messages
Priority Message Problem Solution
None The primary or
secondary recorder is
not connected.
A recording was requested,
but no recorder is available.
Try again, then contact your
technical personnel.
None The primary or
secondary recorder is
out of paper
A recording was requested,
but the recorder is out of
paper.
Replace the recorder paper
(see page 314)
None The primary or
secondary recorder
door is open.
The recorder door is open. Close the door of the recorder.
None The timed recording is
started.
The requested recording is
being printed.
Information only - no action
required.
None The continuous
recording is started.
None The timed recording
request is accepted.
The recorder is not available
and the requested recording
is queued or stored for later
printing.
None The continuous
recording request is
accepted.
None The timed recording is
finished.
The requested recording is
printed.
None The continuous
recording is finished.
None The timed recording is
cancelled.
The requested recording
was manually cancelled.
None The continuous
recording is cancelled.
None The recording is not
accepted
The assigned recorder is not
available and the recording
request was ignored.
Contact your technical
personnel to check the recorder
assignment.
None There is a primary or
secondary recorder
failure.
The recording request was
not accepted due to a
recorder hardware failure.
Contact your technical
personnel.
None The excess artefact
recording is cancelled.
The recording request was
not accepted due to an
artefact.
Check the ECG-lead
connections; contact your
technical personnel.