Alarms
92
Instructions for Use Infinity Acute Care System - Monitoring Applications SW VG1
To change the ARR alarm settings
1 Select the Alarms... button on the main-menu
bar.
2 Select the Limits tab (if not already selected).
3 Select the ARR tab on the right side.
4 Select the corresponding setup button in the
Alarm priority column (C) to select the alarm
priority. A crossed-out triangle appears when
alarm monitoring is disabled. The priority for
asystole and ventricular fibrillation events
cannot be changed. The alarm priority ’high’ is
always assigned to these categories.
5 Select the corresponding setup button in the
Rate column (D) to set the rate.
6 Select the corresponding setup button in the
Count column (E) to set the count.
7 Use one of the following settings in the Archive
column (E) to determine what happens in
response to an alarm:
– Off – no event is stored and no recording is
generated.
– Store – stores the event for later review
(see page 97).
– Rec – generates a timed recording.
– Str/Rec – generates a timed recording and
stores the event.
8 Select the desired arrhythmia mode using the
ARR mode buttons (H).