M588-00-880 Issue E
Page 110 © Edwards Limited 2015. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Appendix A2
A2.4.5.9 Smart stop menu
Scroll by pressing UP/DOWN keys. Press ENTER to open a sub menu or CANCEL to exit back to the Set sequences
menu.
Press the UP/DOWN keys to select enable / disable and press ENTER to accept or CANCEL to exit back to the Smart
stop menu.
A2.4.5.10 Speed control options menu
Scroll by pressing UP/DOWN keys. Press ENTER to open a sub menu or CANCEL to exit back to the Set sequences
menu.
A2.4.5.11 Reset MicroTIM
Empties the MicroTIM configuration set storage of any configurations that have been loaded there.
A2.4.5.12 Default Reset
Resets all of the non volatile configuration settings on the pump (e.g.: delays, temperature setpoints and thresholds)
to their default factory-setting values for that pump type.
Table A16 - Smart stop options menu
Option Text on PDT line 2
Options on PDT line 2 after pressing Enter (use
up/down arrow to change, enter to select)
Smart stop enable/disable Enable/Disable… Enable/Disable
Time for pump to stop Stop time… Range 0 – 3600 seconds in 60 second steps
*
*
Scroll through the options by pressing the UP/DOWN keys. Press ENTER to accept the new setting. Press
CANCEL to exit back into the Smart stop menu.
Size of speed step changes Step size… Range 20 to 110 Hz
*
Temperature to stop pump once
reached
Setpoint… Range 0-210 °C
*
Time after temperature is
reached before pump stops
Settle time… Range 1-200 seconds
*
Table A17 - Speed control options menu
Option Text on PDT line 2
Options on PDT line 2 after pressing Enter (use up/down
arrow to change, enter to select)
Set speed control dry pump Dry pump… Select from:
Normal/PID
External Voltage
SIM protocol
Second speed
Set speed control booster Booster… Select from:
Normal
External Voltage
SIM protocol
Second speed
Set the second speed Second Speed… Range 20 - 110 Hz
*
*
Scroll through the options by pressing the UP/DOWN keys. Press ENTER to accept the new setting. Press
CANCEL to exit back into the Speed control menu.