© Edwards Limited 2015. All rights reserved. Page 131
Edwards and the Edwards logo are trademarks of Edwards Limited.
Appendix A2
M588-00-880 Issue E
A2.12 How to set up the Ethernet port
The GXS pump has a single 10baseT Ethernet port that can be configured in either static address mode or DHCP mode.
There are a variety of protocols that can be run over the Ethernet port including HTTP Web pages, Fabworks
EtherNim, SIM and E54 Modbus.
The web server and SIM protocols are the only protocols available for users. The web server is enabled automatically
when the Ethernet port is operational. The manual for SIM protocol is P411-00-200. If you wish to use the Ethernet
port with different protocols then contact Edwards.
Before the Ethernet connection is used it requires a valid IP address. Set the IP address using the PDT. Both static IP
addressing and DHCP dynamic addressing are supported by the pumps.
Note: If the GXS pump is set to DHCP mode it will not be visible through a router. If operation through a router
is required then use static address mode. The GXS pump does not require NTP, DNS or SMTP server addresses
to be set to function correctly. A gateway address is only required if working with a router. If working in
static address mode then you need to provide the IP address, the address mask and the gateway address.
The following menus are used by the PDT to configure the Ethernet port: