Appendix A3
A3.2 Warnings
The pump controller generates a warning when a problem is encountered.
By default, the GXS pump is prevented from going on-process when there are active warnings so it is important to
investigate the cause of a warning. It is possible to configure the pump so that it will go on-process with active
warnings - refer to A2.6.7 for more information.
Once the problem that caused a warning has been resolved, the warning is cleared by the pump controller.
A3.2.1 LED warning indicators
If the pump encounters a problem, warnings are indicated on the LEDs on the front control panel, the rear panel and
on the PDT if fitted.
The warning LEDs on the front control panel and rear panel illuminate continuously when a warning is generated.
If a PDT is fitted, the warning LED flashes to indicate a new warning. Refer to Appendix A2.3 for more information
on how warnings are indicated and how they can be acknowledged using the PDT.
Once all the warnings are cleared, the warning LEDs extinguish.
A3.2.2 PDT warnings
If a PDT is fitted, each warning triggers a warning message to be displayed. Refer to Appendix A2.3 for more
information on how warnings are handled by the PDT.
Table A29 lists the warning messages that might be displayed on the PDT with possible causes and actions that should
be taken.
EVENT 331.42
Can not start Command.
Pump is not warmed up.
The pump must be warmed up
and in Green Mode/Standby
mode to initiate a DP Clean
sequence.
Either reduce the warm up temperature (refer to
Appendix A2.6.1), enable pump on-process when
cold (refer to Appendix A2.6.2) or wait for the
pump to warm up.
Make sure pump is in Green Mode/Standby mode
before attempting a DP Clean.
EVENT 331.43
Can not start Command.
Pump is not in Green Mode
mode.
The pump must be in Green
Mode/Standby mode to
initiate a DP Clean sequence.
Set the pump into Green Mode/Standby mode and
then start DP clean.
EVENT 331.46
Can not start Command.
Pump is not running.
The pump must be running to
initiate a DP Clean sequence.
Table A29 - Warnings
Warning message on
PDT
‘Action’ message on
PDT
Possible cause Action
Warning 1.01
Power interrupt
Check pwr supply There has been a brownout
of the electrical supply to the
pump lasting more than 1
second.
Check the electrical supply.
Table A28 - Events (continued)
Event message on PDT Possible cause Action