© Edwards Limited 2015. All rights reserved. Page 145
Edwards and the Edwards logo are trademarks of Edwards Limited.
Appendix A3
M588-00-880 Issue E
Warning 39.11
Exh Press High
Exhaust Blocked
Service Pump 
The pressure in the exhaust 
pipeline is too high. 
- A valve in the pipeline may 
be shut. 
- There may be process 
debris or condensation in the 
exhaust pipeline.
-You may have too many 
pumping systems connected 
to the exhaust pipeline.
Refer to Section 4.6 to 
determine the exhaust pressure 
that triggered the warning.
Check that all valves in the 
exhaust line are open and 
consider whether process debris 
or condensation are likely.
Warning 39.13
Sensor missing
- The exhaust pressure 
transducer is not fitted, or is 
disconnected or has failed. 
Check and rectify as necessary.
Warning 51.13
Sup Missing
See manual There is a communication 
problem within the pump 
controller. Refer to 
Appendix A3.5.1.
Cycle the power to the pump. If 
the warning persists, contact 
Edwards to replace the pump 
controller.
Warning 54.11
MB Temp High
See manual The temperature of the 
booster pump is too high. 
- Check that cooling water is 
connected, switched on and is to 
specification given in 
Section 2.6.
Warning 54.13
Sensor missing
- The booster temperature 
sensor may have become 
disconnected or failed.
- Check that the booster pump 
temperature sensor is fitted and 
is correctly connected.
- Check the operation of the 
sensor and replace it if it has 
failed.
Warning 55.11
DP Temp High
Or
Warning 63.11
DP Temp High
See manual The temperature of the 
pump is too high. 
Check that cooling water is 
connected, switched on and is to 
specification given in 
Section 2.6.
Warning 55.13
Sensor missing
Or
Warning 63.13
Sensor missing
- The dry pump temperature 
sensor may have become 
disconnected or failed.
- Check that the dry pump 
temperature sensor is fitted and 
is correctly connected.
- Check the operation of the 
sensor and replace it if it has 
failed.
Warning 71.13
AC Sup Missing
See manual The pump controller cannot 
communicate with the 
accessory module.
- Check that the accessory 
module is correctly connected to 
the pump and then cycle the 
power to the pump.
- If this warning persists, contact 
Edwards.
Warning 152.01
Valve Not Shut
Check ISOL Valve The inlet isolation valve has 
failed to close. 
Check the wiring and air supply 
to the inlet isolation valve. 
NOTE: this warning will only 
clear when the valve has 
successfully been opened and 
closed.
Table A29 - Warnings (continued)
Warning message on 
PDT
‘Action’ message on 
PDT
Possible cause Action