Appendix A3
A3.4 Inverter warnings and alarms
The inverters used to drive the dry pump and booster can also generate warnings and alarms if problems are
encountered. Inverter warning and alarm codes are displayed by the PDT if fitted.
The alarm and warning codes are each reported as a 16-bit word encoded as 4 hexadecimal numbers. The codes are
displayed as follows:
1234 5678
Where:
The first 4 digits signify an alarm code and,
The second 4 digits signify a warning code
Digits 1 – 8 are reported as hexadecimal digits from 0 to F.
To determine the cause of a warning or alarm you must first decode each hexadecimal number. 0 indicates that there
is no alert for that bit. Normally you will only see codes of 1, 2, 4 or 8 for each bit but if two or more events occur
in the same bit then these will be added together.
For example:
5 = 1 + 4 so if code 5 is displayed it means that alerts 1 AND 4 are active
B = 1 + 2 + 8 so if code B is displayed it means that alerts 1, 2 AND 8 are active.
To decode inverter warning and alarm codes, refer to Table A31 to convert the hexadecimal digits into alert
combinations and then refer to Table A32 and A33).
Alarm 184.10
Dry Pump Stopped
See manual The dry pump speed is very
low.
The rotor might be locked.
Cycle the power to the pump and
attempt to re-start. If this fails,
contact Edwards.
Alarm 186.01
DP INV xxxx yyyy
xxxx yyyy
aaaaaaaaaaaaaaaa
Dry pump Inverter has
raised an alarm and
stopped the system.
Refer to Appendix A3.4 for more
information.
Alarm 186.01
DP Not Running
See manual Dry pump Inverter will not
start up when requested.
Check inverter fault history.
Alarm 186.13
No DP Inv Comms
See manual The pump controller cannot
communicate with the dry
pump inverter during start-
up checks.
Check the wiring between the
pump controller and the dry
pump inverter.
Alarm 196.01
DP Inv xxxx yyyy
Diag aaaa bbbb
/zzzzzzzzzzzzzzz
DP 2nd Dry Pump Inverter
has raised an alarm and
stopped the dry pumping
system.
Refer to Appendix A3.4 for more
information.
Alarm 196.13
No DP Inv Comms
See manual The pump controller cannot
communicate with the dry
pump 2nd inverter.
Check the wiring between the
pump controller and the dry
pump inverter.
Alarm 314.12
DP Speed too low
Do not restart
Contact Service
The pump is running more
slowly than the speed that
is demanded.
Contact Edwards. Do not attempt
to start the pump until it has
been checked - this alarm is
latched and the pump will not
restart.