Technical data
M588-00-880 Issue E
Table 9 - Electrical connections
Description
Mating connector description / external
supply rating
Internal supply rating
Mains connection Refer to Table 7 for the mains connector fitted
to each variant. Connector is either:
Harting Han® K4/4-F finger safe 09 38 008
2703, 6-16mm
2
fine stranded wire (VDE 0295
class 5, refer to Table 10 ), 8.9mm max
insulation diameter
or
Harting Han® Axial Screw module 100A (2
off required), part number of mating half
suitable for 10 – 25mm
2
wire is 0914 002 2753 or
0914 002 2751 for 16 –
35mm
2
wire. Use fine stranded wire (VDE 0295
class 5, refer to Table 10).
or
Harting Han® Axial Screw module 200A (3
off required), part number of mating half
suitable for 25 – 40mm
2
wire is 0914 001 2763 or
0914 001 2762 for 40 –
70mm
2
wire. Use fine stranded wire (VDE 0295
class 5, refer to Table 10).
Refer to installation section for wiring
diagram
PDT interface XLR type 5-way plug 24 V d.c. 0.2 A
Pin 1 0 V
Pin 2 24 V
Pin 3 Transmit data
Pin 4 Receive data
Pin 5 Not used
System interface XLR type 5-way plug 24 V d.c. 0.75 A
*
Pin 1 0 V
Pin 2 24 V
Pin 3 Transmit data
Pin 4 Receive data
Pin 5 Not used
Ethernet interface Standard RJ45 type or Neutrik
®
EtherCon
®
RJ45 (IEEE802.3i
10 Base T Ethernet)
EMS interface XLR type 6-way plug
External emergency stop switch
Pin 1 - supply, Pin 2 - return
Note: If there is no external
connection a link plug must be
fitted to operate the pump.
24 V d.c. 100 mA
Internal emergency stop switch
Pin 3 - common, Pin 4 - normally open
30 V a.c. 1 A, 60 V d.c. 0.55 A
Comms 24 V supply
Pin 5 - supply, Pin 6 - 0 V supply common
24 V d.c. 0.75 A
*