EasyManua.ls Logo

Edwards SCU-1601

Edwards SCU-1601
173 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCU-1601 Control Unit for Turbomolecular Pump
MT-96E-1B1-0
Page 96
5
SERIAL COMMUNICATION PROTOCOL
5.4 Command specifications
5.4.1 Command list
Function
code
Command/Query Name
Function
?
D
ReadMeas
Reads the measured rotational speed.
Sp
E
Command
Sends commands START, STOP, RESET (valid
when MANUAL/REMOTE switch is selected ON.)
?
F
ReadFailMess
Reads the errors being detected.
?
M
ReadModFonct
Reads the pump operation mode and the errors
being detected.
?
V
ReadVersion
Reads the software version of the SCU-1601.
?
c
ReadCounters
Reads serial number, running time and start
counter.
?
d
ReadSetPoint
Reads the setting values of the speed set point
and the built-in TMS temperature.
?
e
ReadMotorTemp
Reads the measured motor temperature.
?
f
ReadStatus
Reads the various settings.
(Remote mode, TMS function, INHIBIT and
optional signal [emergency vent valve]).
?
g
ReadEvents
Reads the error record.
Sp
h
SetSpeedSetPoint
Changes the speed set point.
*1
?
h
ReadSpeedSetPoint
Reads the speed set point.
?
m
ReadModFonctWithWarning
Reads the pump operation mode, the errors and
the warnings being detected
?
[
ReadMeasValue
Reads the TMS temperature, motor temperature
and measured rotational speed.
Sp
0
SetOptions
Changes optional function (Second speed
option)
*1
?
0
ReadOptions
Reads optional function (Second speed option,
External TMS temperature)
Table 22 - Command list
*1 There is an endurance limit of setting changes (no more than 24 times per day for about ten years). If this limit is exceeded,
it not only may become impossible to change the setting but can also cause pump failure. Do not make setting changes in
excess of this endurance limit.

Table of Contents

Related product manuals