3.11.9 Warning function setting
Refer to Section 8.2, ""WARNING" message function" for the detail of the "WARNING" function.
1. Warning Damage Point
Enable/Disable Damage limit of the bearing warning function.
When setting to "ENABLE" and the Damage limit of the bearing exceeds the setting value, a warning
message is displayed.
2. Warning Imbalance
Enable/Disable Imbalance warning function.
When setting to "ENABLE" and the Imbalance of the rotor exceeds the setting value, a warning
message is displayed.
3. Warning Pump Runtime
Enable/Disable warning function of the pump operation hours.
When setting to "ENABLE" and the operating hours exceeds the setting value, the warning message
is displayed.
4. Pump Runtime Set Point
Enable/Disable warning function of the pump operating hours in units of 100 hours. (up to 30 million
hours)
5. Warning Pump Overload
Enable/Disable overload warning function of the pump.
When setting "ENABLE" and the state for the motor current of the STP pump continues exceeding or
dropping below the setting value, the warning message is displayed.
6. Current Ceiling Rate
Sets the motor current of the overload warning function of the pump.
The rated speed (max. current) is as 100%. It can be set from 0 to 100 %. The smaller current setting
value is given "WARNING" because the warning occurs when the pump overloads and the motor
current exceeds the setting value.
7. Speed Floor Rate
Sets the rotational speed of the overload warning function of the pump.
The rated speed is as 100%. It can be set from 0 to 100 %. The bigger rated rotational speed setting
value is given "WARNING" because the warning occurs when the pump overloads and the rated
speed drops below the setting value in NORMAL state.