SCU-1601 Control Unit for Turbomolecular Pump
MT-96E-1B1-0
Page 21
3
INSTALLATION
3.5 Attaching the SCU-1601 to a rack
CAUTION
The SCU-1601 cannot be supported with only the 4 screws on the front panel. Always support it from
the bottom using a support angle. When installing the SCU-1601, DO NOT block the ventilation port
of the SCU-1601 by the support angle or other things. It will cause the SCU-1601 to overheat.
The dimensions of the SCU-1601 front panel conform to EIA standards. Therefore, this panel can be
attached to any type of commercially available rack. Attach the SCU-1601 as follows:
Note: For the dimensions of the front panel and positions of the screw holes for the rubber feet, refer
to Figure 11 and Figure 15.
Attach the front panel to the rack using the 4 screws Figure 15 (1).
Support the SCU-1601 from the bottom using a support angle Figure 15 (2) or a similar tool.
When attaching the SCU-1601 to a movable rack, to protect the SCU-1601 during transport,
remove the rubber feet from the bottom and attach the SCU-1601 to the rack using the screw
holes for the rubber feet.
1
2
1.
Front panel 4 securing screws
2.
Mounting support
Figure 15 - Example of securing the SCU-1601