SCU-1601 Control Unit for Turbomolecular Pump
MT-96E-1B1-0
Page 19
3
INSTALLATION
3.4 Precautions before installation
CAUTION
DO NOT move the STP pump and the STP control unit while the STP pump is in operation.
Doing so may result in product damage.
CAUTION
Fasten the STP control unit to a rack to prevent a falling accident caused by earthquake.
3.4.1 Operating environment
The SCU-1601 should be installed in an area which meets the following requirements.
Be sure the ambient temperature of the control unit during STP pump operation:
Ambient temperature
0 to 40 ºC (32 to 104 ºF)
Ambient relative humidity
30 to 95 % (non condensing)
Environment
A place at an altitude of 2,000 m or less
A place indoors
Others: An area free of exposure to direct sunlight, high
humidity, dust, salty air, dripping water, explosive or flammable
gas, corrosive gas, radiation, strong magnetic and electric
fields, excessive vibration and sources of electric noise
Installation condition
Install and anchor the SCU-1601 horizontally (within ±10 degrees)
3.4.2 Insulation test
CAUTION
The varistor for the power supply line protection is installed in the SCU-1601. DO NOT perform the
insulation test with the varistor installed. Doing so may result in product damage.
DO NOT perform an insulation test on the SCU-1601. When performing the insulation test on your
equipment, ensure that you disconnect the SCU-1601 from the equipment that is to be insulation
tested, so that the test voltage is not applied to the SCU-1601.