3.11 Parameter set mode
Parameter set mode is used to set the different parameters of the STP pump and SCU-1601.
3.11.1 Remote operation setting
The remote parallel port (I/O REMOTE), serial port COM1, serial port COM2, and STP-Link are fitted
on the SCU-1601 as standard connectors for the remote operation. The hardware is selected for the
remote operation setting of the start, stop and reset. Refer to Section 4.5 and Section 5 for the remote
operating methods. The remote output monitoring the operating state can be output with any
hardware.
3.11.2 Rotational speed setting
The rotational speed can be changed in units of 500 rpm in the NORMAL state. The variable range is
from half of the rated speed to the rated speed. The rated speed is different according to the
connected pump.
3.11.3 TMS setting
Configures TMS control function (ENABLE
*
/
External/DISABLE).
Set to "External" for the STP-XA2703BV155 / XA3203BV155 pump. The "ENABLE" setting is not
used.
3.11.4 Rotational inhibit signal setting
Enable/Disable the rotation inhibit signal.
3.11.5 Second speed option setting
Enable/Disable the second speed option. Second speed option and option signal setting cannot be
enabled simultaneously.
3.11.6 Second speed setting
The second speed can be changed in units of 500 rpm. The variable range is from half of the rated
speed to the rated speed. The rated speed is different according to the connected pump.
* This function is not used with the STP-XA2703BV155 / XA3203BV155 pump.