SCU-1601 Control Unit for Turbomolecular Pump
SERIAL COMMUNICATION PROTOCOL
5.4.15 ReadMeasValue
Function: Reads the Built-in TMS temperature, motor temperature and measured rotational speed.
The motor temperature value is the same temperature as "ReadMotorTemp". The measured
rotational speed value is the same as "ReadMeas" parameter 2 (Measured rotational speed).
Transmission frame:
*1 The HEX code of ASCII character ’[’ is "5B".
120-bits hexadecimal
coded ASCII
Built-in TMS temperature (Unit ºC)
16-bits hexadecimal coded
ASCII
Motor temperature (Unit ºC)
16-bits hexadecimal coded
ASCII
40-bits hexadecimal coded
ASCII
Measured rotational speed (Unit: Hz)
16-bits hexadecimal coded
ASCII
64-bits hexadecimal coded
ASCII