STP-iS2207 Series Turbomolecular Pump 
 
MT-89E-001-C
Page 132 
7
SAFETY FUNCTIONS 
7.1.8  Overspeed 
When the rotational speed of the STP pump exceeds 38,000 rpm due to a failure in the motor driver, 
the STP pump changes into free-running and stops. 
The  "FAILURE"  LED  (red)  illuminates,  and  a  failure  signal  is  output  from  the  "X2  REMOTE" 
connector. 
 
7.2  "WARNING" function 
7.2.1  "WARNING" function 
The STP pump is provided with a WARNING function when an overhaul is needed following a self-test 
as shown in Table 34. 
The type of warning is indicated by flashing pattern of the "FAILURE" LED (orange). 
When two or more warnings are detected simultaneously, a high-priority warning is indicated. Also, 
the warning signal is output from the "X2 REMOTE" connector. 
The  serial communication,  the  STP-Link (optional  accessory), or  the  display  unit iDT-001 (optional 
accessory) can confirm the error message, and change the settings. 
Refer  to  Section  5,  "Serial  Communication  Protocol",  the  STP-Link  or  the  display  unit  iDT-001 
Instruction Manual of the optional accessories for details. 
No
Warning 
"FAILURE" LED 
indication 
Priority 
1 
Other Warning 
(Other warnings) 
Steady orange  High 
2 
First Damage Limit 
Second Damage Limit 
(Touch down bearing warning function) 
3 flashes in orange   
3 
Pump Run Time Over 
(Pump runtime warning function) 
2 flashes in orange 
 
4 
Pump Overload 
(Pump overload warning function) 
 
5 
Imbalance 
(Excessive imbalance warning function) 
1 flash in orange  Low 
Table 34 - WARNING function