STP-iS2207 Series Turbomolecular Pump 
 
MT-89E-001-C
Page 151
STORAGE AND DISPOSAL 
9
9  STORAGE AND DISPOSAL 
9.1  Storage of the STP pump 
When planning not  to use  the STP pump over a  long period  (more  than  a few  months), follow  the 
precautions below: 
1.  Store the STP pump in a vertical position.  
2.  Close the inlet port of the STP pump and vacuum it using a backing pump. 
3.  Introduce dry N
2
 gas or dry air from the outlet port. 
4.  Close the outlet port. 
5.  If the STP pump is water cooled, introduce compressed air from one side of the cooling water 
port so that no water remains in the STP pump. 
6.  DO NOT store the STP pump in the following areas: 
  Areas  of  high  humidity,  (if  it  must  be  stored  in  a  place  of  high  humidity,  insulate  it  from  the 
outside and use a dehumidifying agent.) 
  Areas prone to temperature extremes. High temperatures of no more than +55 °C (131 °F), and 
low temperatures of less than –25 °C (-13 °F). 
  Areas where corrosive gases may exist. 
  Areas subjected to water/dampness. 
  Areas subjected to excessive dust. 
  Areas with insufficient ventilation. 
  Areas subjected to strong magnetic and electric fields or radiation. 
  Areas which will be subjected to direct sunlight. 
  Areas subjected to mist. 
  Areas subjected to electric noise and vibration.