STP-iS2207 Series Turbomolecular Pump 
 
MT-89E-001-C
Page 136 
7
SAFETY FUNCTIONS 
7.4  WARNING function setting 
The WARNING functions can be set to Enable or Disable. Set to "ENABLE" when using the 
WARNING function. Set to "DISABLE" to release each WARNING function after the WARNING is 
detected. 
A setting value of the "Pump Run Time Over" and "Pump Overload" are adjustable. The WARNING 
detection can be released by setting the value larger than pump running hours after "Pump Run Time 
Over" is detected. Also, the WARNING detection can be released by powering off the pump after 
"Pump Overload" is detected.   
See Table 35 and Table 36 for the default setting. The setting can be changed via serial 
communication, the STP-Link (optional accessory) or the display unit iDT-001 (optional accessory). 
WARNING function  Default setting  User setting 
User setting 
change (threshold)
Second Damage Limit  Enable  Disable  Disable 
First Damage Limit  Enable  Enable  Disable 
Pump Run Time Over  Disable  Enable  Enable 
Pump Overload  Disable  Enable  Enable 
Imbalance  Enable  Enable  Disable 
Other Warning  Enable  Disable  Disable 
Table 35 - Default setting of WARNING function 
WARNING function 
Factory setting 
(threshold) 
Variable range
User setting 
change (threshold) 
Pump Run Time Over  0 hour 
0 to 100,000 
hours 
It can be set in units of 100
hour. 0 is not counted. 
Pump Overload 
Motor 
current 
100.0 %
0 to 100.0 % 
The ratio (%) to the rated 
current value. Smaller 
value has high possibility to 
get "WARNING". 
 
Rotational 
speed 
0.0 %  0 to 100.0 % 
The rate (%) to the rated 
speed value. Larger value 
has high possibility to get 
"WARNING". 
Table 36 - Factory Setting and Variable Range 
Note:  Refer to Section 5, "Serial Communication Protocol", STP-Link, or the display unit iDT-001 
Instruction Manuals for the setting methods.