STP-iS2207 Series Turbomolecular Pump 
 
MT-89E-001-C
Page 134 
7
SAFETY FUNCTIONS 
7.3  Contents of WARNING function 
7.3.1  First Damage limit 
Impact of the STP pump rotor onto the touch-down bearing, such as by an unexpected in-rush of air 
from outside or in the event of power failure, can damage the touch-down bearings. The STP pump 
monitors these impacts and assigns damage points to the event of "Disturbance" or "Power Failure". 
When the accumulated damage point attains certain points, the "First Damage Limit" is detected. 
"First  Damage  Limit"  warning  can  be  released  by  setting  the  warning  function  to  disable  after  it  is 
detected.  The  accumulated  damage  points  can  be  confirmed  and  the  setting  can  be  changed  via 
serial  communication,  the  STP-Link  (optional  accessory)  or  the  display  unit  iDT-001  (optional 
accessory).   
 
7.3.2  Second Damage Limit 
"Second Damage Limit" is detected when continuing to operate the STP pump after "First Damage 
Limit"  is  detected  and  the  accumulation  of  the  damage  point  attains  the  setting  value  by  further 
"Disturbance" and "Power Failure". 
"Second Damage Limit" warning detects "START NOT ALLOWED" failure at the same time. 
Overhaul  the  STP  pump  because  the  STP  pump  cannot  perform  the  rotational  operation  when 
"Second Damage Limit" is displayed. 
When  operating  the  STP  pump  is  needed  after  "Second  Damage  Limit"  is  displayed,  disable  the 
"Second  Damage  Limit". The  rotating  operation can  be performed even  though  the  "WARNING" is 
detected. 
The  setting  can  be  changed  via  serial  communication,  the  STP-Link  (optional  accessory)  or  the 
display unit iDT-001 (optional accessory). 
 
7.3.3  Pump run time over 
There is a run-time counter that tracks the total running hours of the pump. When the run-time counter 
attains the setting value, the "Pump Run Time Over" is detected. When deposit accumulates inside 
the STP pump, this function can be used to consider the time of the overhaul. 
The setting value is user-definable. 
Confirmation of the total running hours of the pump and change of the setting value can be performed 
via  serial  communication,  the  STP-Link  (optional  accessory)  or  the  display  unit  iDT-001  (optional 
accessory).