TECHNICAL DATA 
2
2.3  Water cooling use condition 
Use cooling water which fulfills the following conditions. 
Item  Specification 
Port type 
Rc 1/4 (Female)
Flow rate 
L/min
3 
Water temperature  °C
15 to 35 
*1
 
Water pressure  MPa  (kgf/cm
2
)
0.3 (3) 
Maximum grain size  mm
2
0.03 
pH 
6.5 to 8.0 
Water hardness  mg/L
100 
Resistivity 
kΩ・cm
4 to 1000 
Turbidity  FNU
< 30 