MAINTENANCE AND INSPECTION 
8
8.4.3  Recommended maintenance intervals for main parts 
Internal components of the STP pump will be damaged from gas load, heat, and corrosion when 
used long-term. Deterioration or abrasion of the internal parts of the STP pump will cause 
unexpected failures. When overhauling the STP pump, replacement of main parts are recommended. 
Refer to the following list as recommended maintenance intervals for main parts. 
Part name 
Recommended maintenance intervals 
*1
 
Corrosive gas 
Inert gas or 
vacuum pumping only 
Rotor blades  3 years  5 years 
Rotor shaft  3 years  5 years 
Motor, sensor, electromagnet, 
connector 
5 years  7 years 
Air-cooling fan  5 years 
PCBA in the control unit  7 years 
Accessories
TMS valve  3 years 
  TMS heater  3 years 
  Connection cable  5 years 
*1: Maintenance intervals of any parts may be shortened depending on the condition of use. 
 
The costs of replacing parts that need to be replaced when deterioration or abrasion is observed will 
be at the customers charge, even though a period is not exceeding the recommended maintenance 
intervals. 
When overhaul of the STP pump is needed, contact Service office.