SAFETY FUNCTIONS 
7.6.2  Indication of "FAILURE" LED (green) 
If an abnormality/error is detected by self test during the power ON operation, "FAILURE" LED flashes 
in green. This is the indication of STP pump or control unit failure. Contact Service Office. 
 
7.6.3  No Indication of the "FAILURE" LED 
No.
Symptom  Probable cause  Countermeasures 
Referred 
Section 
1  The "POWER" 
LED does not 
illuminate. 
Incorrect connection of 
power cable 
Connect the power cable correctly.  3.4.5 
3.4.6 
Abnormal power voltage
Set the input voltage to the 
±10 %. 
2.4 
Power failure  Check whether or not a power failure 
has occurred. 
7.1.1 
 
  Insufficient cooling  Make sure that the STP pump is cooled.
2.3 
2  The 
"ROTATION" 
LED does not 
flash after 
performing the 
start operation. 
Failure of the "Input 
Operation Port" setting. 
Set the input operation port correctly.  4.9 
3  Insufficient 
ultimate 
pressure. 
Failure of the 
backing-pump. 
Check the capacity and starting state 
(use a backing-pump having more 
capacity that we specified). 
3.5 
    Leakage of the piping 
system. 
Check the piping system for leakage.  3.5 
    Residual molecules.  If the main composition is H
2
 or H
2
O, 
perform baking; if it is other gases, clean 
the inside of the vacuum equipment (If 
gases remain inside the STP pump, 
contact Service office when it is needed 
to be cleaned). 
4.3 
4  Abnormal noise 
is generated 
while the pump 
is rotating 
External vibration or 
impact 
Remove external vibration so as not to 
transmit it to the STP pump. 
3.1.1 
Table 43 - Troubleshooting with no indication of the "FAILURE" LED