STP-iS2207 Series Turbomolecular Pump 
 
MT-89E-001-C
Page 42 
INSTALLATION 
3
3.4.6  How to connect the power cable 
WARNING 
 
When connecting/disconnecting cables, always power off the primary power (switch 
the MAIN POWER to "OFF") and isolate (Lockout/Tagout) the electrical energy 
source, water and gas, and other energy sources on the vacuum equipment. Failure 
to do so may result in the inadvertent rotation of the STP pump which may result in 
an accident, an electric shock or damage to equipment. Moreover, an accident 
caused by water leaks or gas leak may occur. 
 
CAUTION 
 
Use the power cable that Edwards has specified. The use of other cables may result in product 
damage. 
 
CAUTION 
 
Confirm the power voltage on the label. Connect the power cable securely to prevent incorrect 
wiring. 
 
CAUTION 
 
Connect each cable securely with caution, avoiding any obstacles. 
DO NOT place heavy objects on the cables or bend them excessively. If any problem occurs in 
cables, connectors or terminals, the STP pump may not function normally. 
DO NOT apply voltage to each connector pin and DO NOT cause any short-circuiting between 
pins. 
Install cables so that personnel are not exposed to risk of tripping or falling. 
 
CAUTION 
 
With each connector align the position of the guide key of the connectors and insert vertically so as 
not to bend the pins. If a pin is bent, not only may the connector not function normally, but it may 
make the pins contact, resulting in a malfunction. Lock and securely tighten each connector and 
screw. 
 
CAUTION 
 
DO NOT apply surge voltage exceeding 1 kV to the input power line. Always ground the power 
cable to prevent electric shock.