STP-iS2207 Series Turbomolecular Pump
MT-89E-001-C
Page 116
5
SERIAL COMMUNICATION PROTOCOL
Parameter
13
ASCII
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
HEX
*1 Assign Fhex or the reading data of ReadOptionFunc (?=).
5.4.18 ReadCondition
Function: Reads the pump model and damage point.
Transmission frame:
*1
PCSIM
Stx
0
0
1
?
{ Etx
LRC
SIMPC
Ack
PCSIM
*1
Ack
SIMPC
Stx
0
0
1
Sp
{ Parameter 1 to 4 Etx
LRC
*1 The HEX code of ASCII character ' { ' is "7B".
Parameter
Item Data format Remarks
1 Pump model 160-bits hexadecimal
coded ASCII
2 [System reservation] 32-bits hexadecimal
coded ASCII
3 Damage point 16-bits hexadecimal
coded ASCII
4 [System reservation] 64-bits hexadecimal
coded ASCII