STP-iS2207 Series Turbomolecular Pump 
 
MT-89E-001-C
Page 28 
INSTALLATION 
3
3.3.2  Name and function of the control unit 
Figure 17 shows the front panel of the control unit.  
WARNING 
 
A hazardous live voltage may exist at the connectors that are marked with the  
warning sign 
 
. DO NOT touch the connecter. Doing so may result in electric 
shock. When connecting/disconnecting the connecter, always power off the primary 
power (switch the MAIN POWER to "OFF") and isolate (Lockout/Tagout) the 
electrical energy source, water and gas, and other energy sources on the vacuum 
equipment. 
 
CAUTION 
 
The "STP-LINK" connector (X5) is waterproof and dust-proof when the cover is attached. When 
the STP-Link and Display unit iDT-001 are connected, the waterproof and dust-proof state of the 
pump cannot be maintained. DO NOT leave the STP-Link and Display unit iDT-001 constantly 
connected to the pump in an environment affected by dust and water splashes. There is a risk that 
the STP pump may fail.