SERIAL COMMUNICATION PROTOCOL 
  DO NOT bundle a communication line with a protective earth conductor or a power line. 
Moreover, keep away a communication line from the apparatus used as a noise source as much 
as possible. 
  As radio frequency noise measure, place a ferrite core on both ends of the communication cable. 
When electromagnetic interference caused by radio frequency noise in frequency band (150 kHz 
to 1 GHz) affects communication, attaching ring ferrite cores to the cable is effective to reduce 
communication failure. 
 
Figure 39 - Example of ring ferrite core installation 
  Locate and secure the cables. 
It may be difficult to measure the reproducibility of the communication failure without securing the 
cables. 
  Avoid installing a power line and a communication line in the same metallic duct. 
When  unavoidable,  separate  a  line  with  a  metal  separator  certainly,  and  connect  the  duct 
containing  a  metal  separator  certainly  to  GND,  or  installing  a  communication  line  put  into 
conductive pipes, such as metal. 
             
 
Communication line
  Metallic separator
GND
Power line 
Metallic duct
 
Figure 40 - Example of cable installation in metallic duct