Troubleshooting multi EA 5000
108
13.4 Displays in the window
Status analyzer
13.4.1 Overview
In the
Status analyzer
window, information on the device status or information on indi-
vidual modules is displayed.
TN liquid vertical (1) - liquid
Status analyzer multi EA 5000
Rack Syringe: 112 - : 50 µL
C-NDIR
N-CLD-5000
S-UVFD-5000
MFC 1
MFC 2
MFC 3
Furnace temperature
Cl-POT
OK
0.15
300
0
100
1050 °C
1
2
3
4
5
6
Figure65Status analyzer window
1 Window name and basic module 2 Active method and sample state
3 Sampling modules 4 Detectors with sensor values
5 Gas flow displays 6 Furnace temperature (horizontal
line – horizontal installation, vertical
line – vertical installation)
The displays in the
Status analyzer
window are color-coded. The colors have the follow-
ing meanings:
Color Description
Black Status of the corresponding component OK, device ready for measure-
ment
Gray Detector is inactive
Green Detector OK, device ready for measurement (OK)
Or
Detector busy, measurement can only be started once the routine is
complete (detector-specific)
Red Component not ready for measurement:
¡ Run-in period not yet complete; wait for run-in time to complete
¡ Error: Troubleshooting, access the corresponding information for
the component via the
System
|
Component test
menu item.