multi EA 5000 Troubleshooting
107
Error
code
Error message/cause Remedy
411
415
420
Syringe pump: Pump step not permitted
Syringe pump: Command error
Syringe pump: incorrect type
Communication error Acknowledge message
Initialize
13.3 Initializing the basic module and the system components
Initialization of a measuring system establishes communication between the measuring
system and the computer. The multiWin program differentiates between a standard ini-
tialization and a basic initialization.
With standard initialization, only the system components active before the last shut-
down of multiWin are queried and the method last active is loaded.
The basic initialization, on the other hand, is more thorough and tests all connected sys-
tem components activated in the multiWin program in the window. The basic initializa-
tion must always be performed in the following situations:
¡ Connection of new system components
¡ Recognition of system components which were shut down or were not connected
during the last initialization
¡ Fault in the communication between the measuring system and the computer
Performing the basic initializa-
tion
Basic initialization is always performed if the
Device - edit
window was opened and ex-
ited via
[OK]
:
} Select the
Device
|
Device - edit
menu item.
} Make any necessary changes and exit the
Device - edit
window via
[OK]
.
} Click on
[Initialize analyzer]
in the main window.
ü The system is initialized and the method last used is activated. If the initialization
was successful, the
[Start Measurement]
,
[Activate method]
and possibly the
[Start calibration]
buttons are displayed in the main window.
Standard initialization
Click on the
[Initialize analyzer]
button in the main window. Alternatively, select the
Sys-
tem
|
Initialize
menu item.