multi EA 5000 Chlorine analysis with the Cl module
53
The "high sensitive" measuring
cell
This measuring cell is used for very low chlorine content (0.01 to 10 µg). In particular, it
is recommended for use in EOX determination or for gaseous sample.
The measuring cell consists of the base element with magnetic stirring rod and lid. The
positions of the required components are marked inside the lid of the measuring cell.
The central borehole is intended for an adapter that is connected to the exhaust hose. It
is used as an exhaust for the acetic acid vapors.
The small unmarked opening is used for direct injection into the cell. This opening is
closed with a plug when measuring gas is routed into the measuring cell from the com-
bustion furnace.
Figure30Measuring cell "high sensitive" with lid
1 Measuring cell base element with
magnetic stirrer rod
2 Opening for the sensor electrode
3 Opening for the silver electrode 4 Opening for the platinum electrode
5 Opening for the gas inlet tube 6 Opening for direct injection
7 Opening for exhaust connection