EasyManua.ls Logo

Endress+Hauser Analytik Jena multi EA 5000 - Figure 28 Combined Electrode; Figure 29 Measuring Cell High Concentration with Lid

Endress+Hauser Analytik Jena multi EA 5000
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chlorine analysis with the Cl module multi EA 5000
52
1
2 3 4
5
Figure28Combined electrode
1 Connector plug 2 Generator cathode (Pt)
3 Indicator electrodes (Ag) 4 Gas inlet to measuring cell
5 Measuring gas connection
(hose20)
The "high concentration" mea-
suring cell
The measuring cell has the same function as the "sensitive" measuring cell, but has a
higher electrolyte volume. It is suitable for a chlorine content of 10 to 1000 µg
, and is
recommended in particular for TX determination of highly contaminated waste and of
polymer samples with high PVC content.
Here, as well, the maintenance-free combined electrode is used.
1 3 4 5 62
Figure29Measuring cell "high concentration" with lid
1 Knurled head screw 2 Opening for combined electrode
(labeled)
3 Opening for direct injection and
exhaust connection
4 Electrode space with silver anode
5 Magnetic stirring rod 6 Electrical connection of the
measuring cell