Chlorine analysis with the Cl module multi EA 5000
52
Figure28Combined electrode
1 Connector plug 2 Generator cathode (Pt)
3 Indicator electrodes (Ag) 4 Gas inlet to measuring cell
5 Measuring gas connection
(hose20)
The "high concentration" mea-
suring cell
The measuring cell has the same function as the "sensitive" measuring cell, but has a
higher electrolyte volume. It is suitable for a chlorine content of 10 to 1000 µg
, and is
recommended in particular for TX determination of highly contaminated waste and of
polymer samples with high PVC content.
Here, as well, the maintenance-free combined electrode is used.
Figure29Measuring cell "high concentration" with lid
1 Knurled head screw 2 Opening for combined electrode
(labeled)
3 Opening for direct injection and
exhaust connection
4 Electrode space with silver anode
5 Magnetic stirring rod 6 Electrical connection of the
measuring cell